Containment-oriented cascade
The pressure relationship directs leakage inward, with the interchange maintained below the work chamber.
Unlike the positive-pressure model, this enclosure uses a negative pressure arrangement for hazardous-drug containment. The interchange is maintained more negative than the main chamber, while filtered air recirculation reduces the external exhaust requirement compared with a total-exhaust isolator.
The pressure relationship directs leakage inward, with the interchange maintained below the work chamber.
HEPA-filtered recirculation supports ISO Class 5 conditions without exhausting the entire chamber flow.
A polycarbonate glove window provides process visibility while separating the operator from the preparation.
| Nominal Width | 1.2 m |
|---|---|
| Work Zone | 895 x 591 x 695 mm |
| Pressure Mode | Negative-pressure recirculating |
| Interchange Pressure | More negative than the main work chamber |
| Air Changes | 20 per minute |
| Air Cleanliness | ISO Class 5 HEPA-filtered work zone |
| Construction | Stainless steel enclosure |
| Glove Window | Polycarbonate barrier panel |
| Intended Compounds | Sterile hazardous drugs |
| Tùy chỉnh | Exhaust interface, interchange side and work-zone services can be tailored to the hazardous compounding process. |
Sterile hazardous-drug preparation in pharmacies that need negative-pressure containment while limiting the volume sent to facility exhaust.
Interchange orientation, base stand, work-surface accessories and facility exhaust connection can be selected for the pharmacy layout.