No room recirculation
All cabinet airflow leaves through the hard-duct exhaust route.
This compact B2 model sends all cabinet air to the facility exhaust rather than recirculating it to the room. The 1379 mm exterior width provides a focused work zone for suitable volatile or hazardous procedures while limiting the footprint and exhaust demand relative to the six-foot unit.
All cabinet airflow leaves through the hard-duct exhaust route.
A dedicated 254 mm connection supports facility system planning.
The four-foot format suits focused processes that do not need the 1800 mm work width.
| Exterior Dimensions | 1379 x 803 x 1816 mm |
|---|---|
| Cabinet Class | Class II Type B2 |
| Exhaust Architecture | 100% total exhaust |
| Exhaust Connection | 254 mm |
| Inflow Velocity | Approximately 0.53 m/s |
| Downflow Velocity | Approximately 0.28 m/s |
| Facility Connection | Hard duct |
| Work-Zone Use | Focused volatile or hazardous process |
| Nominal Size | 1200 mm / 4 ft |
| Tùy chỉnh | Duct transition, services and base can be configured for the facility exhaust and process. |
Suitable biological work involving limited volatile chemicals or hazardous-drug processes requiring total cabinet exhaust.
Duct transition, services, base stand and exhaust monitoring interface can be configured.
Gửi cho chúng tôi kích thước của bạn, điều kiện lắp đặt, mục tiêu hiệu suất và bề mặt yêu cầu hoàn thiện.