No room recirculation
All cabinet airflow leaves through the hard-duct exhaust route.
This compact B2 model sends all cabinet air to the facility exhaust rather than recirculating it to the room. The 1379 mm exterior width provides a focused work zone for suitable volatile or hazardous procedures while limiting the footprint and exhaust demand relative to the six-foot unit.
All cabinet airflow leaves through the hard-duct exhaust route.
A dedicated 254 mm connection supports facility system planning.
The four-foot format suits focused processes that do not need the 1800 mm work width.
| Exterior Dimensions | 1379 x 803 x 1816 mm |
|---|---|
| Cabinet Class | Class II Type B2 |
| Exhaust Architecture | 100% total exhaust |
| Exhaust Connection | 254 mm |
| Inflow Velocity | Approximately 0.53 m/s |
| Downflow Velocity | Approximately 0.28 m/s |
| Facility Connection | Hard duct |
| Work-Zone Use | Focused volatile or hazardous process |
| Nominal Size | 1200 mm / 4 ft |
| Penyesuaian | Duct transition, services and base can be configured for the facility exhaust and process. |
Suitable biological work involving limited volatile chemicals or hazardous-drug processes requiring total cabinet exhaust.
Duct transition, services, base stand and exhaust monitoring interface can be configured.