General exhaust construction
The weather-resistant galvanized enclosure supports ducted final filtration without the added bagging system.
This standard exhaust housing uses a 14-gauge galvanized cabinet, gasket-seal HEPA track and access from either side. An optional nominal 2 in prefilter protects the final element, and the family pressure capability reaches plus or minus 8 in. w.g. Unlike the BIBO version, this model is intended for exhaust streams whose hazard assessment permits conventional side-door filter removal.
The weather-resistant galvanized enclosure supports ducted final filtration without the added bagging system.
Doors on both sides allow filter withdrawal from the clear mechanical-room aisle.
A shallow prefilter option reduces coarse dust accumulation on the exhaust HEPA.
| Housing Material | 14-gauge galvanized steel |
|---|---|
| Access | Either side |
| Filter Type | Gasket-seal HEPA |
| Prefilter Track | 2 in nominal optional |
| Family Pressure Capability | ±8 in. w.g. |
| Change Method | Conventional side-door removal |
| Cabinet Duty | Weather-resistant exhaust housing |
| Filter Clamping | Mechanical gasket compression |
| Installation | Non-containment ducted exhaust |
| การปรับแต่ง | สามารถใช้ได้ตามแอปพลิเคชันที่ตั้งใจไว้และสภาพแวดล้อมการทํางาน |
For general laboratory, medical and clean-process exhaust systems where HEPA filtration is required but the risk assessment does not require a contained change bag.
Confirm hazard classification, cabinet material, HEPA depth, gasket side, prefilter, access side, weather protection, test ports and duct flanges.
ส่งมิติของคุณเงื่อนไขการติดตั้งเป้าหมายประสิทธิภาพและผิวที่ต้องการ