Pemasok One-Stop Anda untuk Peralatan Cleanroom dan Fasilitas Aseptik EsensialDukungan proyek internasional
Open mobile navigationClose mobile navigation
Beranda / Produk / Containment & Biosafety Equipment / Aseptic Processing Isolator 1150 mm Work-Zone Two-Glove
STJH-ASEPTIC-PROCESSING-ISOLATOR-1150-2G

Aseptic Processing Isolator 1150 mm Work-Zone Two-Glove

An intermediate two-glove aseptic isolator with an 1150 mm chamber and more staging width than the compact platform.
  • 1150 mm work zone
  • Two-glove process layout
  • 1060 W published load
Permintaan untuk kutipan Browse Produk

Perkenalan produk

The intermediate model adds 210 mm of work-zone width without adding a third glove port. It is suited to processes that need broader component staging around a single two-glove manipulation position, with published 1060 W consumption and a 355 kg gross equipment weight.

Fitur utama:

Expanded side staging

The extra chamber width holds more components beside the central manipulation area.

Familiar two-glove ergonomics

The model increases capacity without changing the operator to a three-position handling pattern.

Dedicated exhaust stream

All process air is discharged rather than returned through a recirculation loop.

Parameter teknis

External Dimensions1650 x 810 x 1580 mm including airlock
Work-Zone Dimensions1150 x 600 x 660 mm
Airlock Dimensions350 x 400 x 435 mm
Glove Access2 ports
CleanlinessISO Class 3
Air Velocity0.40 m/s ± 0.05 m/s
Ventilation100% exhaust
PressurizationPositive-pressure operation
Published Utilities1060 W; 355 kg gross weight
Customization (Kustomisasi)Process services, airlock position, glove system and controls can be adapted to the sterile handling sequence.

Applications

Medium-throughput aseptic assembly, sterile formulation and preparation workflows that need additional staging within a two-glove barrier.

Solutions for Your Application

The 940 mm chamber leaves components stacked near the glove working area.Use the 1150 mm span to reserve lateral staging space while retaining a two-glove operation.
A recirculating loop is not wanted for the selected process ventilation strategy.Use the documented full-exhaust architecture and size the facility connection for the operating mode.

Pilihan yang tersedia

Airlock side, pressure controls, glove ports, utilities, monitoring sensors and work-surface fittings are configurable.

Planning a Aseptic Processing Isolator 1150 mm Work-Zone Two-Glove specification?

Kirimkan kami dimensi Anda, kondisi instalasi, target kinerja dan permukaan akhir yang diperlukan.