Expanded side staging
The extra chamber width holds more components beside the central manipulation area.
The intermediate model adds 210 mm of work-zone width without adding a third glove port. It is suited to processes that need broader component staging around a single two-glove manipulation position, with published 1060 W consumption and a 355 kg gross equipment weight.
The extra chamber width holds more components beside the central manipulation area.
The model increases capacity without changing the operator to a three-position handling pattern.
All process air is discharged rather than returned through a recirculation loop.
| External Dimensions | 1650 x 810 x 1580 mm including airlock |
|---|---|
| Work-Zone Dimensions | 1150 x 600 x 660 mm |
| Airlock Dimensions | 350 x 400 x 435 mm |
| Glove Access | 2 ports |
| Cleanliness | ISO Class 3 |
| Air Velocity | 0.40 m/s ± 0.05 m/s |
| Ventilation | 100% exhaust |
| Pressurization | Positive-pressure operation |
| Published Utilities | 1060 W; 355 kg gross weight |
| التخصيص | Process services, airlock position, glove system and controls can be adapted to the sterile handling sequence. |
Medium-throughput aseptic assembly, sterile formulation and preparation workflows that need additional staging within a two-glove barrier.
Airlock side, pressure controls, glove ports, utilities, monitoring sensors and work-surface fittings are configurable.
أرسل لنا الأبعاد الخاصة بك، شروط التثبيت، أهداف الأداء والتشطيب السطحي المطلوب.