Pembekal Satu Henti Anda untuk Peralatan Bilik Bersih dan Kemudahan AseptikSokongan Projek Antarabangsa
Open mobile navigationClose mobile navigation
Laman Utama / Produk / Containment & Biosafety Equipment / Aseptic Processing Isolator 1150 mm Work-Zone Two-Glove
STJH-ASEPTIC-PROCESSING-ISOLATOR-1150-2G

Aseptic Processing Isolator 1150 mm Work-Zone Two-Glove

An intermediate two-glove aseptic isolator with an 1150 mm chamber and more staging width than the compact platform.
  • 1150 mm work zone
  • Two-glove process layout
  • 1060 W published load
Permintaan sebut harga Semak imbas produk

Pengenalan Produk

The intermediate model adds 210 mm of work-zone width without adding a third glove port. It is suited to processes that need broader component staging around a single two-glove manipulation position, with published 1060 W consumption and a 355 kg gross equipment weight.

Ciri-ciri Utama

Expanded side staging

The extra chamber width holds more components beside the central manipulation area.

Familiar two-glove ergonomics

The model increases capacity without changing the operator to a three-position handling pattern.

Dedicated exhaust stream

All process air is discharged rather than returned through a recirculation loop.

Parameter teknikal

External Dimensions1650 x 810 x 1580 mm including airlock
Work-Zone Dimensions1150 x 600 x 660 mm
Airlock Dimensions350 x 400 x 435 mm
Glove Access2 ports
CleanlinessISO Class 3
Air Velocity0.40 m/s ± 0.05 m/s
Ventilation100% exhaust
PressurizationPositive-pressure operation
Published Utilities1060 W; 355 kg gross weight
PenyesuaianProcess services, airlock position, glove system and controls can be adapted to the sterile handling sequence.

Applications

Medium-throughput aseptic assembly, sterile formulation and preparation workflows that need additional staging within a two-glove barrier.

Solutions for Your Application

The 940 mm chamber leaves components stacked near the glove working area.Use the 1150 mm span to reserve lateral staging space while retaining a two-glove operation.
A recirculating loop is not wanted for the selected process ventilation strategy.Use the documented full-exhaust architecture and size the facility connection for the operating mode.

Pilihan yang boleh didapati

Airlock side, pressure controls, glove ports, utilities, monitoring sensors and work-surface fittings are configurable.

Planning a Aseptic Processing Isolator 1150 mm Work-Zone Two-Glove specification?

Hantarkan kepada kami dimensi anda, syarat pemasangan, sasaran prestasi dan kemasan permukaan yang diperlukan.

Berbincang dengan pasukan kami
Email Kami WhatsApp WeChat

Laman Utama

Produk

Berita

Tentang Kami

Hubungi Kami