Focused testing layout
Two glove ports place essential test handling within one compact operator reach zone.
A 1200 x 540 x 700 mm internal chamber provides a defined station for limited sample volumes and an integrated sterility-test pump. The two glove positions suit a focused workflow, while the compact chamber requires less process air than the 1600 and 2000 mm versions.
Two glove ports place essential test handling within one compact operator reach zone.
U15 supply filtration supports the clean process environment, with H14 filtration on the exhaust path.
The chamber can be arranged for total exhaust or a lower-volume recirculating operating mode.
| Internal Process Chamber | 1200 x 540 x 700 mm |
|---|---|
| External Process-Chamber Envelope | 1504 x 881 x 2430 mm |
| Glove Ports | 2 |
| Supply Filter | U15, nominal 99.9995% at MPPS |
| Exhaust Filter | H14, nominal 99.995% at MPPS |
| Total-Exhaust Airflow | 1021 m³/h |
| Recirculating Airflow | 510 m³/h |
| Typical Process Equipment | Integrated sterility-test pump |
| Relative Family Size | Smallest standard chamber in the retained range |
| 사용자 정의 | Transfer chamber, pump arrangement and utility connections can be tailored to the sterility-test method. |
Pharmaceutical quality-control sterility testing with limited sample volume, compact equipment staging and two-glove manipulation.
Pass chambers, sterility-test pump integration, monitoring instruments and exhaust configuration can be arranged around the laboratory method.
치수, 설치 조건, 성능 목표 및 필요한 표면 마감을 보내십시오.