Controlled process surface
The dedicated zone supports temperature-sensitive sample preparation inside the protected cabinet.
This variant integrates a local controlled surface area into the B2 work zone. An external circulator supplies the approximately 270 x 300 mm zone for temperature-sensitive handling while the cabinet retains its total-exhaust airflow architecture.
The dedicated zone supports temperature-sensitive sample preparation inside the protected cabinet.
A compatible circulator supplies controlled fluid through the work-surface system.
Cabinet air is exhausted to the facility rather than returned to the laboratory.
| Nominal Cabinet Width | 1219 mm |
|---|---|
| Cabinet Class | Class II Type B2 |
| Controlled Work Area | Approximately 270 x 300 mm |
| Temperature Range | 0–50 °C with external circulator |
| Circulator Flow | 1.5–1.9 L/min |
| Exhaust Architecture | 100% total exhaust |
| Facility Connection | Hard duct |
| Service Features | Cable portals and instrument service fixture |
| Primary Use | Temperature-sensitive contained handling |
| Customization (Kustomisasi) | Controlled-zone location, circulator connections and cabinet services can be configured for the process. |
Temperature-sensitive biological sample handling, cooled preparation and instrument work requiring Type B2 total exhaust.
Controlled-zone position, circulator connection, service penetrations and base stand can be configured.