Votre fournisseur à guichet unique pour les équipements de salle blanche et les essentiels des installations aseptiquesSoutien international aux projets
Open mobile navigationClose mobile navigation
Accueil / Produits / Équipement Cleanroom / Laboratory Specimen Pass Box 305 mm Turntable
STJH-LABORATORY-SPECIMEN-PASS-BOX-TURNTABLE-305

Laboratory Specimen Pass Box 305 mm Turntable

A rotating-cylinder specimen transfer that preserves a privacy barrier without two opposing swing doors.
  • 305 mm rotating cylinder
  • 180-degree manual transfer
  • Removable cylinder for sterilization
Demander un devis. Browse les produits

Introduction de produit

This recessed unit uses a 305 mm-diameter stainless cylinder instead of a conventional cubic chamber. Staff place the sealed container in the 167 x 152 mm opening and rotate the cylinder 180 degrees for rear retrieval. Self-lubricating stainless bearing plates support movement, bullet catches control rotation, and a tumbler lock releases the cylinder for sterilizing or autoclaving.

Key Features

Continuous visual barrier

The curved cylinder blocks a direct sightline between patient and laboratory sides during transfer.

Controlled rotation

Positioned catches resist uncontrolled movement while stainless bearing plates support smooth manual operation.

Removable transfer element

A keyed release allows the cylinder to be removed when separate sterilization or autoclaving is required.

Paramètres techniques

Overall Face375 W x 394 H mm
Cylinder Diameter305 mm
Transfer Opening167 W x 152 H mm
Partition Opening337 W x 356 H mm
Supported Partition Depth86-178 mm
ConstructionType 304 satin stainless steel
Transfer MotionManual 180-degree rotation
Bearing SystemSelf-lubricating stainless ball-bearing plates
Cylinder RetentionBullet catches with tumbler-lock removal
CustomisationDisponible en fonction de l'application prévue et de l'environnement d'exploitation.

Applications

Designed for sealed urine cups, tissue containers and small clinical samples transferred from a patient or collection area to laboratory staff. It is not intended for trays larger than the 167 x 152 mm transfer opening.

Solutions for Your Application

Opposing doors can reveal the laboratory side or allow direct reach-through.The rotating cylinder maintains a physical privacy barrier throughout the 180-degree handoff.
A fixed transfer interior is difficult to sterilize away from the wall.Unlock and remove the cylinder for the defined sterilization or autoclave process.

Options disponibles

Confirm partition depth, face trim, lock keying and collection-side labeling. Cleaning procedures should match the stainless finish and removable cylinder handling method.

Planning a Laboratory Specimen Pass Box 305 mm Turntable specification?

Envoyez-nous vos dimensions, conditions d'installation, objectifs de performance et finition de surface requise.