Continuous visual barrier
The curved cylinder blocks a direct sightline between patient and laboratory sides during transfer.
This recessed unit uses a 305 mm-diameter stainless cylinder instead of a conventional cubic chamber. Staff place the sealed container in the 167 x 152 mm opening and rotate the cylinder 180 degrees for rear retrieval. Self-lubricating stainless bearing plates support movement, bullet catches control rotation, and a tumbler lock releases the cylinder for sterilizing or autoclaving.
The curved cylinder blocks a direct sightline between patient and laboratory sides during transfer.
Positioned catches resist uncontrolled movement while stainless bearing plates support smooth manual operation.
A keyed release allows the cylinder to be removed when separate sterilization or autoclaving is required.
| Overall Face | 375 W x 394 H mm |
|---|---|
| Cylinder Diameter | 305 mm |
| Transfer Opening | 167 W x 152 H mm |
| Partition Opening | 337 W x 356 H mm |
| Supported Partition Depth | 86-178 mm |
| Construction | Type 304 satin stainless steel |
| Transfer Motion | Manual 180-degree rotation |
| Bearing System | Self-lubricating stainless ball-bearing plates |
| Cylinder Retention | Bullet catches with tumbler-lock removal |
| Customization | Disponible según la aplicación prevista y el entorno operativo. |
Designed for sealed urine cups, tissue containers and small clinical samples transferred from a patient or collection area to laboratory staff. It is not intended for trays larger than the 167 x 152 mm transfer opening.
Confirm partition depth, face trim, lock keying and collection-side labeling. Cleaning procedures should match the stainless finish and removable cylinder handling method.
Envíenos sus dimensiones, condiciones de instalación, objetivos de rendimiento y acabado superficial requerido.