Compact HPAPI station
The two-glove chamber concentrates weighing and dispensing around one operator position.
This smallest retained high-potency model uses a 1200 mm process chamber and two oval glove ports. Its normal-run exhaust is 20 m³/h, the lowest in the family, while the stainless chamber and double-stage H14 exhaust path support contained weighing within a compact external envelope.
The two-glove chamber concentrates weighing and dispensing around one operator position.
Two push-push H14 filter stages provide a contained service approach on the exhaust path.
The system has a higher published exhaust duty for operation following a defined barrier breach condition.
| External Dimensions with One Pass Chamber | 1990 x 924 x 2275 mm |
|---|---|
| Internal Process Chamber | 1200 x 633 x 842 mm |
| Glove Ports | 2 oval ports, 200 x 300 mm |
| Chamber Pressure | Negative |
| Airflow Pattern | Turbulent flow, more than 30 air changes per hour |
| Normal-Run Exhaust | 20 m³/h at 300 Pa |
| Breach Exhaust | 100 m³/h at 900 Pa |
| Exhaust Filtration | Double-stage H14 push-push, nominal 99.995% at 0.3 µm MPPS |
| Internal Chamber | 4.0 mm 316L stainless steel, Ra ≤ 0.40 µm |
| Tùy chỉnh | Pass chamber, weighing equipment and transfer interfaces can be configured for the potency and batch workflow. |
Compact high-potency active ingredient weighing, dispensing and small-batch sampling behind a two-glove negative-pressure barrier.
Pass-chamber arrangement, integrated scales, transfer valves and contained waste interfaces can be selected for the process.
Gửi cho chúng tôi kích thước của bạn, điều kiện lắp đặt, mục tiêu hiệu suất và bề mặt yêu cầu hoàn thiện.