Method-specific arrangement
The working positions prioritize filter canister handling rather than the broad sample staging of the batch model.
This model keeps the chamber short and dedicates its internal layout to the filtration canister, sample containers and rinse sequence used for membrane-method testing. Integrated hydrogen-peroxide decontamination prepares the closed work zone before the operator begins the two-glove test campaign.
The working positions prioritize filter canister handling rather than the broad sample staging of the batch model.
Hydrogen peroxide is distributed through the enclosure before aeration and aseptic release.
Supply and exhaust HEPA stages support the controlled Grade A or ISO 5 work environment.
| Glove Ports | 2 |
|---|---|
| Primary Test Method | Membrane filtration |
| Work-Zone Classification | EU Grade A / ISO 5 |
| Air Velocity | Approximately 0.45 m/s |
| Supply Filtration | Double HEPA configuration |
| Exhaust Filtration | Double HEPA configuration |
| Decontamination | Integrated H2O2 micro-nebulization cycle |
| Operation | Manual membrane-test workflow |
| Monitoring | Continuous process and environmental data option |
| Customização | Canister layout, glove spacing, transfer interface and monitoring instruments can be configured for the test method. |
Routine sterility testing of filterable parenterals, ophthalmic products and other liquid samples using the membrane-filtration method.
Filtration hardware, transfer hatch, glove system, particle monitoring and sample-rack arrangement can be specified.
Envie-nos as suas dimensões, condições de instalação, metas de desempenho e acabamento de superfície necessário.