Extended process sequence
Four reach positions can support weighing, subdivision, packaging and output within one negative-pressure barrier.
This model provides the maximum standard process width in the family. Four glove ports span a 2000 mm chamber, allowing equipment, incoming material and discharge steps to be separated along the enclosure; normal-run exhaust rises to 40 m³/h for the larger volume.
Four reach positions can support weighing, subdivision, packaging and output within one negative-pressure barrier.
The 2790 mm width with a pass chamber accommodates equipment that cannot fit the shorter models.
Double H14 push-push filter stages support a planned contained replacement approach.
| External Dimensions with One Pass Chamber | 2790 x 924 x 2275 mm |
|---|---|
| Internal Process Chamber | 2000 x 633 x 842 mm |
| Glove Ports | 4 oval ports, 200 x 300 mm |
| Chamber Pressure | Negative |
| Airflow Pattern | Turbulent flow, more than 30 air changes per hour |
| Normal-Run Exhaust | 40 m³/h at 300 Pa |
| Breach Exhaust | 100 m³/h at 900 Pa |
| Exhaust Filtration | Double-stage H14 push-push, nominal 99.995% at 0.3 µm MPPS |
| Internal Chamber | 4.0 mm 316L stainless steel, Ra ≤ 0.40 µm |
| Customization | Process equipment, transfer ports and chamber services can be configured for the extended high-potency workflow. |
Large-batch HPAPI weighing, multi-step dispensing and extended contained preparation requiring four glove-working positions.
Multiple transfer devices, integrated weighing equipment, wash accessories and process-service penetrations can be arranged for the full sequence.
Send us your dimensions, installation conditions, performance targets and required surface finish.