Product Introduction

The eight-foot model is intended for centralized production where multiple contained positions must remain inside one negative-pressure barrier. Its long work surface can support staged ingredient receipt, preparation, verification and pack-off, while one or two transfer chambers define how materials enter and leave.

Key Features

Extended production line

The largest footprint accommodates several sequential operations without moving open work outside the barrier.

Multi-user access planning

Two- or four-port architecture can be selected around reach zones and division of operator duties.

Contained exhaust maintenance

Bag-out filtration supports service planning for high-use hazardous-drug production.

Technical Parameters

Standard Length8 ft
Depth2.5 ft
Height6.75 ft
Main ChamberNegative pressure, ISO Class 4, vertical unidirectional airflow
Transfer ChambersOne or two depending on material-flow design
Transfer Air ClassISO Class 5 at static and dynamic conditions
AirflowAt least 400 cfm; at least 40 fpm at 6 in from filter face
FiltrationHEPA, nominal 99.99% DOP at 0.3 µm, with bag-out
Glove SystemTwo- or four-port architecture with 10 in stainless shoulder rings
CustomizationOperator stations, transfer direction, disposal interfaces, stands, utilities and internal fixtures can be configured for the intended central-pharmacy workload.

Applications

Centralized oncology production, high-volume hazardous sterile preparation and multi-stage compounding that requires several contained work positions.

Solutions for Your Application

A large daily batch requires multiple stages but open handoff between cabinets is undesirable.Arrange receipt, manipulation, checking and pack-off along the continuous eight-foot chamber.
The largest unit may obstruct staff circulation or service access.Confirm the full equipment envelope, door swings and filter-maintenance side against the room layout before selecting this length.

Available Options

Dual interchanges, four-port glove layout, adjustable lifting system, secondary filtration, waste and sharps interfaces, IV bars and electrical outlets are available.