Compact unidirectional zone
The small chamber maintains a downward clean-air path around laboratory-size items.
The smallest vertical-flow model provides a 500 mm cubic work zone beneath a 365 x 365 mm HEPA filter. A nominal 0.45 m/s downward airflow sweeps particles toward the return path while using only 0.2 kW connected power. Its 750 x 500 x 900 mm outer envelope suits small wall openings, but payloads must remain below the supply diffuser and leave airflow clearance.
The small chamber maintains a downward clean-air path around laboratory-size items.
A 0.2 kW rating suits installations with limited local electrical capacity.
The 365 mm filter matches the compact work zone and reduces replacement-media size.
| Useful Dimensions | 500 W x 500 D x 500 H mm |
|---|---|
| Outside Dimensions | 750 W x 500 D x 900 H mm |
| HEPA Filter Size | 365 x 365 mm |
| Air Velocity | 0.45 m/s nominal |
| Connected Power | 0.2 kW |
| Airflow Direction | Vertical downward unidirectional flow |
| Door Arrangement | Opposing interlocked doors |
| Construction | Powder-coated steel or SUS304 |
| Typical Load | Small sterile components and packs |
| Customisierung | Verfügbar je nach beabsichtigter Anwendung und Betriebsumgebung. |
For sterile samples, compact medical components and laboratory material transfer where active vertical wash is desired. Keep packages low enough to avoid disturbing the diffuser and validate air velocity with the normal load in place.
Select powder-coated or SUS304 construction, interlock type, pressure gauge and purge timer. Coordinate the 750 mm equipment width with service access.
Senden Sie uns Ihre Abmessungen, Installationsbedingungen, Leistungsziele und gewünschte Oberflächenveredelung.