موردك الموحد لمعدات الغرف النظيفة وأساسيات المرافق العقليةدعم المشاريع الدولية
Open mobile navigationClose mobile navigation
الرئيسية / المنتجات / معدات الغرف النظيفة / Vertical Laminar Flow Clean Bench 2 ft
STJH-VERTICAL-CLEAN-BENCH-2FT

Vertical Laminar Flow Clean Bench 2 ft

The smallest vertical clean bench width for non-hazardous sample preparation in a limited laboratory footprint.
  • 2 ft compact work width
  • 45-65 fpm vertical downflow
  • ISO Class 5 work zone
طلب عرض أسعار تصفح المنتجات

مقدمة المنتج

This 2 ft bench directs HEPA-filtered air downward over a compact work zone. A nominal 45-65 fpm downflow, 99.99% efficient HEPA filter at 0.3 µm and aerodynamic sash opening support ISO Class 5 conditions for non-hazardous work. Its narrow width is best for one compact instrument or a tightly defined manual step rather than broad assembly layouts.

الخصائص الرئيسية

Small laboratory footprint

The two-foot width fits short benches and dedicated sample-preparation corners.

Product-focused downflow

HEPA-filtered air descends across the work surface to reduce particulate deposition on non-hazardous materials.

Adjustable fan output

A variable-speed blower supports velocity setup as filter resistance changes over service life.

المعلمات الفنية

Nominal Width2 ft
Airflow DirectionVertical downflow
Downflow Velocity45-65 fpm
Final FilterHEPA, 99.99% at 0.3 µm
Work-Zone ConditionISO Class 5
Sash Opening9.4 in nominal
Sash Material6.35 mm tempered safety glass
Blower ControlVariable speed
Utility AccessTwo side utility ports
التخصيصمتوفر على أساس التطبيق المقصود وبيئة التشغيل.

Applications

For PCR setup, media plating and small sterile non-hazardous preparation tasks. This clean bench protects the work, not the operator, and must not be used for biohazards, toxins or radionuclides.

Solutions for Your Application

A standard four-foot bench consumes too much scarce laboratory counter space.Use the two-foot model for one compact instrument and a narrowly defined clean task.
Operators may mistake a clean bench for personnel containment.Limit the unit to non-hazardous work and select a containment cabinet where operator protection is required.

الخيارات المتاحة

Choose electrical configuration, airflow monitor, UV lamp package, stand and work-surface material. Utility port fittings can be matched to the process connection.

Planning a Vertical Laminar Flow Clean Bench 2 ft specification?

أرسل لنا الأبعاد الخاصة بك، شروط التثبيت، أهداف الأداء والتشطيب السطحي المطلوب.

تحدث مع فريقنا
E البريد الإلكتروني لنا WhatsApp WeChat