Ductless installation
The built-in blower returns filtered air to the room without a remote exhaust connection.
This two-foot system combines a built-in blower and single HEPA filter in a compact recirculating enclosure. It requires no external duct for normal room-return operation, making it suitable for limited bench space and small potent-powder quantities. Application suitability depends on the compound risk and whether single-stage recirculation is acceptable.
The built-in blower returns filtered air to the room without a remote exhaust connection.
The two-foot work zone concentrates capture around one small balance operation.
Blower and filter remain within the enclosure system for straightforward placement.
| Nominal Width | 2 ft |
|---|---|
| Pressure Mode | Negative work zone |
| Filter Stages | Single HEPA |
| Blower | Built-in |
| Exhaust Mode | Room recirculation |
| External Ducting | Not required |
| Work Capacity | One compact balance |
| Application | Small potent-powder weighing |
| Monitoring | Airflow indication by configuration |
| التخصيص | متوفر على أساس التطبيق المقصود وبيئة التشغيل. |
For small potent-powder weighing where bench space and duct access are limited. A qualified risk assessment must confirm that single-HEPA room recirculation is appropriate.
Select work-surface material, sash, airflow monitor, electrical supply and filter service accessories. More demanding compounds may require double filtration or an isolator.