Routine load capacity
The chamber accepts medium carriers that cannot fit the 457 mm model.
This 24-inch cube balances useful load space with standard wall mounting. The larger volume accommodates medium supply packs and sample carriers while the recirculating HEPA system maintains an ISO 5 internal airflow condition.
The chamber accepts medium carriers that cannot fit the 457 mm model.
A nominal 99.99% at 0.3 µm filter cleans the closed chamber loop.
Equal width, depth and height simplify wall-opening and shelf planning.
| Internal Chamber | 610 x 610 x 610 mm |
|---|---|
| Air Cleanliness | ISO 5 internal airflow |
| Final Filter | HEPA, nominal 99.99% at 0.3 µm |
| Airflow | Closed-loop recirculating |
| Construction | 304 stainless steel |
| Door Arrangement | Two-sided pass-through |
| Load Scale | Medium |
| Room Air Exchange | No continuous room exhaust |
| Installation | Standard wall mount |
| การปรับแต่ง | Door orientation, interlock and internal shelf can be tailored to the material route. |
Routine movement of sterile supply packs, sample carriers and medium instruments across cleanroom pressure boundaries.
Door interlocks, viewing windows, shelf and mounting trim can be configured.
ส่งมิติของคุณเงื่อนไขการติดตั้งเป้าหมายประสิทธิภาพและผิวที่ต้องการ