Home / Products / Cleanroom Doors, Windows & Interlocks / Airtight Cleanroom Swing Door Four-Side Inflatable-Seal
zhengfeng-AIRTIGHT-CLEANROOM-SWING-DOOR-INFLATABLE-SEAL

Airtight Cleanroom Swing Door Four-Side Inflatable-Seal

A stainless hinged assembly inflates a four-sided elastomer seal after closure, with monitored air supply and a defined release sequence.
  • Pressure monitoring makes the seal supply condition observable.
  • Four-sided contact includes the bottom of the opening.
  • A flush floor interface supports frequent wheeled movement.
Request a Quote Browse Products

Active sealing after the leaf reaches its closed position

This stainless swing-door configuration uses a four-sided inflatable silicone or EPDM gasket. Compressed air expands the seal against the closed assembly, including its flush floor interface. Releasing the air retracts contact before the door moves. The sequence differs from a fixed gasket that is compressed by closing force and requires pneumatic controls as well as door hardware.

Door position, lock state, seal pressure and supply pressure are different signals. A closed-position switch does not establish seal inflation, and normal seal pressure does not by itself prove that the entire doorway meets its leakage criterion. The control sequence and the physical boundary must both be validated for the intended process.

Reference configuration and pneumatic duties

ItemSpecification basis
Leaf/frame304 or 316 stainless construction with stainless hinges.
SealFour-sided inflatable silicone or EPDM, selected for gap geometry and cleaning exposure.
Air pressureThe listed approximately 2.5 bar option is a reference inflation setting; the selected seal and regulator define the actual value.
MonitoringIncoming supply and seal-pressure monitoring with defined alarm thresholds and delay.
FloorFlush stainless threshold, with continuity at corners and the wall/floor junction.
OptionsFlush glazing, pneumatic release override and backup supply arrangements.

Inflation pressure is not the permitted room pressure difference. Increasing it cannot compensate indefinitely for a misaligned frame or damaged profile. Specify required air quality, supply range, local regulation, exhaust path and connections for the actual package.

Operating sequence and failure behavior

Normal operation should establish the required closed and secured condition before inflation, then prove the specified pressure state. An opening request requires seal release before movement. Define the response to pressure loss, electrical failure, a damaged seal and emergency operation; these depend on the facility's approved access and containment strategy. Backup air has a finite capacity and needs a documented duty rather than an assumed holding time.

Plan isolation and access for replacing the full gasket and control valves. Inspect contact surfaces, tubing and corners, and check alarm settings and release behavior during maintenance. Avoid pulling the door against an inflated seal.

Quotation and acceptance scope

Quote the complete door, subframe, seal, pneumatic and electrical control package, monitoring interfaces, backup provisions, installation and tests. Acceptance includes the approved sequence, pressure alarms, release/override functions and a separate leakage test at the specified room differential. No biosafety classification, decontamination suitability or absolute gas-tightness follows from inflation alone.