Balanced bench width
The three-foot deck accommodates routine vessels and small instruments without the footprint of the 1200 mm cabinet.
This 914 mm version expands the usable deck while retaining a single angled front access area. It draws room air into the enclosure and filters exhaust air, prioritizing operator and environmental protection for tasks that do not require clean downflow over the product.
The three-foot deck accommodates routine vessels and small instruments without the footprint of the 1200 mm cabinet.
A hinged safety-glass sash provides a clear view while defining the controlled front opening.
HEPA filtration captures airborne particulate before exhaust air is returned or conveyed onward.
| Nominal Work Width | 914 mm |
|---|---|
| Cabinet Class | Class I |
| Inflow Velocity | 0.38-0.53 m/s |
| Pressure Mode | Negative relative to the laboratory |
| Exhaust Filter | HEPA, nominal 99.99% efficiency at 0.3 µm |
| Sash Type | Angled hinged safety glass |
| Airflow Arrangement | Inward room air with filtered exhaust |
| Product Protection | Not supplied |
| Operator Capacity | One operator with a moderate equipment footprint |
| 사용자 정의 | Exhaust routing, services and support configuration can be adapted to the intended laboratory application. |
Routine microbiology, sample preparation and small equipment operation where personnel and environmental protection are required but sterile product airflow is unnecessary.
Exhaust-canopy connection, utility fittings and an airflow monitoring package can be discussed for the selected installation.
치수, 설치 조건, 성능 목표 및 필요한 표면 마감을 보내십시오.