Low-fleet entry point
One head can be moved through a defined port sequence when simultaneous testing is not needed.
The single-head configuration is aimed at small barriers and facilities that prefer one mobile instrument rather than a permanently assigned fleet. An inflatable seal closes the selected port, the integrated pump establishes the test pressure, and the head transfers the measured decay result wirelessly for traceable review.
One head can be moved through a defined port sequence when simultaneous testing is not needed.
Automatic seal inflation removes a manual clamping step before the pressure-decay cycle begins.
The rechargeable head communicates wirelessly, avoiding a data cable across the barrier work area.
| Test Capacity | 1 glove port per test cycle |
|---|---|
| Measurement Principle | Pressure decay based on ISO 10648-2 |
| Minimum Detectable Leak | 100 µm class |
| Port Seal | Automatically inflated adapter seal |
| Enclosure Protection | IP54 according to IEC 60529 |
| Inflation Time | Less than 30 seconds in the referenced system |
| Wireless Protocol | ZigBee |
| Port Identification | Automatic RFID recognition |
| Battery Endurance | Up to 20 cycles per charge |
| Customization | Test disc, glove recipe, pressure settings and mobile accessories can be configured for the port design and operating environment. |
Pre-use and routine integrity checks on pharmaceutical isolator and RABS gloves in small production, compounding or quality-control installations.
Port-specific test discs, cleanroom tablet, mobile trolley, additional glove recipes, charging accessories and central software connection may be selected.
Отправьте нам ваши габариты, условия монтажа, целевые показатели производительности и требуемую отделку поверхности.