Ваш универсальный поставщик оборудования для чистых помещений и асептических средствПоддержка международных проектов
Open mobile navigationClose mobile navigation
Главная / Продукция / Containment & Biosafety Equipment / High-Potency Containment Isolator 1600 mm Three-Glove
STJH-HIGH-POTENCY-CONTAINMENT-ISOLATOR-1600-3G

High-Potency Containment Isolator 1600 mm Three-Glove

A medium-width three-glove high-potency isolator providing additional staging space and 30 m³/h normal-run exhaust.
  • 1600 mm working width
  • Three-glove process access
  • Intermediate exhaust duty
Запрос котировки Просмотреть продукты

Введение продукции

The 1600 mm internal chamber adds a third oval glove position for separating material entry, weighing and discharge activities. Its 2390 mm overall width with one pass chamber and 30 m³/h normal exhaust place it between the compact and four-glove models.

Основные характеристики

Three-position workflow

The extra glove port provides a distinct reach zone for staging or transfer beside the primary weighing position.

Smooth stainless chamber

The 4 mm 316L internal shell uses a low-roughness finish suited to repeatable cleaning.

Controlled negative pressure

Turbulent airflow and negative chamber pressure support containment during medium-scale dispensing.

Технические параметры

External Dimensions with One Pass Chamber2390 x 924 x 2275 mm
Internal Process Chamber1600 x 633 x 842 mm
Glove Ports3 oval ports, 200 x 300 mm
Chamber PressureNegative
Airflow PatternTurbulent flow, more than 30 air changes per hour
Normal-Run Exhaust30 m³/h at 300 Pa
Breach Exhaust100 m³/h at 900 Pa
Exhaust FiltrationDouble-stage H14 push-push, nominal 99.995% at 0.3 µm MPPS
Internal Chamber4.0 mm 316L stainless steel, Ra ≤ 0.40 µm
CustomizationGlove layout, pass chamber and dispensing equipment can be adapted to the material flow and batch size.

Applications

Medium-scale HPAPI weighing, potency-controlled dispensing and multi-step sample preparation requiring three manipulation positions.

Solutions for Your Application

Material staging crowds the same reach zone as the balance.Use the third glove position for dedicated staging or transfer while keeping the weighing area clear.
The installation requires a predictable normal exhaust load.Plan the branch for the published 30 m³/h at 300 Pa normal duty and the separate breach-response condition.

Доступные варианты

Balance integration, rapid-transfer ports, pass-chamber layout and contained cleaning accessories may be selected for the formulation.